Chapter 2 Performance Tests
Test Equipment Required
2-4 User’s Guide
Test Equipment Required
Table 2-1. Recommended Test Equipment
Instrument Required Characteristics Recommended Model Use
*
* C= Calibration (Option 010 only)
OV = Operational Verification
P = Performance Tests
Signal Generator Frequency Range: 100 kHz to 10 MHz
Output Level: +6 to +16 dBm
Harmonic Distortion:<-30 dBc
Spurious Distortion:<-80 dBc
HP 8663A (or equivalent) OV, P
Digitizing
Oscilloscope
Frequency range: 100 kHz to 100 MHz HP 54600B (or equivalent) OV
Universal Counter Frequency Range: 100 kHz to 100
MHz: >10 digits resolution @ 10 MHz
HP 53131A (or equivalent) C (for
Option 010
only)
10 MHz Reference
Source
10 MHz Reference Signal
Accuracy:
±
2 X 10
-12
HP 5071A (or equivalent) C
Terminal or computer Communication software, RS-232
connection
Any Model OV, P, C
RMS Voltmeter Accuracy:
±
2% at 3V Full Scale,
1MHz
HP 3400B (or equivalent) P
50 BNC
Feedthrough
_____________________________ HP 10100C (or equivalent) OV, P, C
RS-232 Interface
Cable
DTE-to-DTE 9-pin (f) to 9-pin (f) HP 24542U (or equivalent) C
Type N-to-BNC
Connector
_____________________________ H P 1250-0780 (or
equivalent)
OV, P
50 BNC Coaxial
Cable (2)
RG-58, 12 meter, male-to-male HP 10503A (or equivalent) OV, P, C